Realization of Laryngeal Stroboscopic Effect Using Continuous Light |
Interdisciplinary Program for Cognitive Science, Pusan National Univ. 1, DepT. of Biomedical Engineering, School of Medicine, Pusan National Univ. 2, Dept. of Otorhinolaryngology, Pusan National Univ. Hosp. 3 |
Geunhyo KIM,
Geun-Hyo KIM 13, Hee-June PARK 23, Won-Jae CHA 3,Jeon-Yeob JANG3,Jae-Keun CHO 3, Seong-Chan SHIN 3,Byung-Joo LEE 3,Soon-Bok KWON 1,Soo-Geun WANG 23
|
¸ñÀû: Laryngeal stroboscopeis a gold standard method in evaluation of
vocal folds
lesions. Many ENT clinics could not use due toit's expensive
price. Purpose
of this study is to examine laryngeal stroboscopic effect that
can be
implemented only with the functions of the CCD camera itself
under
continuous illumination, without any specific hardware or
software system. ¹æ¹ý:We made frequency controllable fan motor system connected with
digital
multimeter for checking the stroboscopic effect in any rotational
frequency(Hz) and controlled shutter speed of CCD camera for degree
of
clearing both fan image and mucosal edge of vocal cords. °á°ú:Rotational frequency of 50, 100, 150Hz appears fixed fan motor image which
means fully synchronization of rotational fan motor and CCD camera. If
shutter speed of CCD camera fixed above 1/10,000second, fan image was very
clearly expressed as compared to the other rotational frequency. In
laryngoscopic finding, shutter speed of 1/2000, mucosal wave and vocal fold
edge were most clearly. °á·Ð:We observed the stroboscopic effect using continuous light on a
specific
rotational frequency, it is useful for evaluation of vocal cords
lesions. |
|